Pressure range |
Atmosphere to 10-9 mbar/Torr |
Accuracy * |
Typically ±15% <100 mbar and ±30%<10-3 mbar |
Maximum over pressure |
6 bar absolute (87 psia) |
Power supply |
+14.5 to +36 V d.c. |
Power consumption |
2 W maximum |
Output signal |
1.8 to 10.2 V d.c. |
Adjustments |
Atmosphere and setpoint |
Set point |
Open collector transistor |
Maximum voltage |
40 V d.c. |
Current |
100 mA maximum |
Temperature range |
|
Operating |
+5 to +60 oC |
Storage |
0 to +70 oC |
Materials exposed to vacuum (Both NW and CF versions) |
Stainless steel (AISI 304, 316, 321, 347), Fluoroelastomer, soda lime glass, Tungsten, trace of Nickel and Nickel Iron |
Internal volume |
26 cm3 |
Weight |
0.8 kg |
External interface connector |
8-way FCC68 / RJ45 Socket |
Interface cables |
Use range of active gauge cables |
Standards |
|
Electromagnetic compatibility |
EN 61326 Industrial Location, Class B emissions |
Enclosure rating |
IP40 |
Pin allocation ** |
|
1. Power supply positive |
5. Signal common |
2. Power supply common |
6. Set-point output |
3. Gauge output |
7. Atmosphere calibration |
4. Gauge identification |
8. Not connected |