Mass |
85 g |
Internal volume |
5 cm3 |
Enclosure rating |
IP40 |
Measurement range (APG100-XM) |
Atmosphere to 10-3 mbar |
Measurement range (APG100-XLC) |
Atmosphere to 10-4 mbar |
Accuracy (APG100-XM) |
Typically +/- 15% at <100 mbar |
Accuracy (APG100-XLC) |
Typically +/- 15% at <10 mbar |
Maximum over-pressure |
10 bar absolute |
Operating temp range |
5o to 60o C |
Storage temp range |
-30o to 70 oC |
Bake-out with no electronics |
150 oC |
Humidity |
80% RH up to 31 oC decreasing linearly |
|
to 50% RH at 40 oC and above |
Maximum altitude |
3000 m |
Filament temperature |
100 oC above ambient |
Electrical supply voltage |
15 to 30 V d.c. nominal |
|
13.5 V d.c. minimum |
|
32 V d.c maximum |
Power consumption |
1 W |
Output signal |
0 to 10 V d.c. nominal |
Set-point – open collector transistor |
|
Rating |
30 V d.c. 100 mA |
Range of set-point |
1.8 to 9.2 V dc |
Fixed hysteresis |
500 mV (1/2 decade) |
Level setting resolution |
6 mV |